Best Practices for Blood-Based Amyloid Pathology Testing

Discover how to maintain plasma sample quality through collection, processing, storage, and preparation for reliable analysis of the pTau 217/β-Amyloid 1-42 Plasma Ratio test.

Brought to you byFujirebio
| 1 min read
Register for free to listen to this article
Listen with Speechify
0:00
1:00
Sample collection icon
1. Sample Collection
Collect whole blood samples via venipuncture using a K2EDTA tube.
icon for sample preparation
4. Sample Preparation
Insufficient equilibration to RT or incomplete thawing can result in uneven analyte distribution. Bring refrigerated plasma samples to RT and thaw frozen samples at RT for at least 45 minutes. Before testing, thoroughly mix samples using a vortex mixer and a centrifuge.
icon for sample processing
2. Sample Processing
Samples degrade rapidly at room temperature (RT). To preserve integrity, centrifuge samples within two hours of collection to separate plasma from blood cells and platelets. Once separated, proceed to testing.
icon for measurement
5. Measurement
Immediately dispense the prepared plasma into sample cups and load them onto the analyzer for pTau 217 and β-amyloid 1-42 testing. Run both tests using the same patient sample and the same analyzer.
icon for storage
3. Storage
If testing cannot be performed immediately after centrifugation, transfer plasma to polypropylene tubes and store refrigerated for up to one day or frozen at -60°C or cooler for up to one month. Avoid storage and shipment at RT.
icon for result interpretation
6. Result Interpretation
The pTau 217/β-amyloid 1-42 Plasma Ratio is calculated from the two test results. FDA-cleared thresholds are used to classify the likelihood of amyloid pathology as negative (≤0.00370), indeterminate (0.00371–0.00737), or positive (≥0.00738).