Turning a gene therapy candidate into a medicine is not just a biology problem. It is a measurement problem. When the same adeno-associated virus (AAV) batch produces different titer numbers depending on which laboratory runs the assay, dose justification becomes an argument rather than a data package, and regulators notice.

Adrien Savy, PhD
AAV Principal Scientist
Coave Therapeutics
In this Innovation Spotlight brought to you in partnership with Bio-Rad, Adrien Savy, AAV principal scientist at Coave Therapeutics, explains how running Bio-Rad's droplet digital PCR (ddPCR) platform across research and manufacturing reduces regulatory risk, what the method still cannot reveal about a batch, and how analytical demands shift as programs move toward commercial scale.
Unlike standard curve-dependent methods, ddPCR partitions each sample into tens of thousands of nanoliter-sized droplets and counts positive and negative reactions directly, enabling absolute quantification of vector genome titer without a plasmid reference standard. This reduces lot-to-lot variability in conventional quantitative PCR (qPCR)-based titration.
What does an AAV analytical development lab do in a gene therapy company like Coave?
Coave develops ligand-conjugated AAV vectors through the ALIGATER™ platform, so analytical development here carries a double remit. The first covers the conventional AAV panel: vector genome titer by digital PCR (dPCR) — specifically Bio-Rad’s droplet digital PCR (ddPCR) platform, capsid titer, full, empty, and partial capsid distribution, genome integrity, identity, VP ratio, residual host-cell and plasmid DNA, and potency.
The second covers the attributes that the conjugation chemistry introduces: degree of conjugation, how it is distributed across the capsid population, and linkage stability. These are approached by mass spectrometry alongside orthogonal methods.
Structurally, the team sits between process development and quality. It translates process questions into measurable attributes, decides what is phase-appropriate, qualifies the methods, and hands them over in a form a QC analyst or a contract development and manufacturing organization (CDMO) can run independently.
Why is bridging R&D and GMP analytics so difficult in AAV gene therapy?
There is very little metrology underneath this field. Reference standard material exists for AAV2 and AAV8 only, and the collaborative studies behind them were sobering: the AAV8 standard's genome titer estimate spans close to two orders of magnitude at ±2 SD across participating laboratories, and PCR-based titration in those exercises varied by up to two logs between labs. For an engineered or conjugated capsid, there is no external anchor at all — every lab is its own island.
The nature of the sample can also shift depending on the method applied. Research and development (R&D) methods are built on crude lysates and small-scale batch material; good manufacturing practice (GMP) release methods run on purified drug product formulated in buffer containing surfactant, at higher particle concentration and in different vial types. Matrix, dilution scheme, and freeze-thaw history all move the numbers.
Finally, the two environments want opposite things. R&D optimizes for flexibility; GMP optimizes for lockdown. Methods are typically qualified and locked before the process has completed its development phase.
What does running different analytical platforms across R&D and QC cost a gene therapy company?
Three costs, in ascending order of pain.
The visible one is bridging work. Every platform difference requires a formal comparability exercise: retained samples, both methods run in parallel, and a statistical equivalence assessment. Each exercise takes an analytical scientist several months, and a full AAV panel requires several methods.
The second is the correction factor problem. Quantitative PCR (qPCR) and ddPCR do not necessarily agree, and the gap is not constant. Using a supercoiled rather than linearized plasmid standard shifts qPCR titer roughly two-fold on its own, and inverted terminal repeat (ITR)-targeted assays overestimate relative to internal targets. If the first-in-human dose was set on the R&D platform and QC releases on another, the team is either restating the dose in the Investigational Medicinal Product Dossier (IMPD) or defending a correction factor at a scientific advice meeting.
The third, and the one people most often underestimate, is investigational blindness. When a release result trends out of range and there is no comparable historical dataset, it is impossible to tell whether the batch moved or the method did. Process comparability becomes confounded, because the process and the measurement changed together.
Same platform is not the same result.
—Adrien Savy, Coave Therapeutics
How does using the same ddPCR platform for R&D and GMP quality control reduce regulatory risk?
Using the same platform removes a whole category of argument. If the toxicology lot, the engineering runs, and every GMP batch were quantified on the same partitioning chemistry, one continuous titer history runs from dose selection through to release. Dose justification becomes traceable rather than reconstructed.
It also fits the direction ICH Q14 and Q2(R2) have taken — both adopted by the FDA in March 2024 and in force in the European Union (EU) since June 2024. Both guidelines are built around an analytical target profile and a lifecycle in which development data becomes part of the validation evidence. One platform makes that story coherent instead of piecemeal.
Absolute quantification helps too: no standard curve means no plasmid standard lot to bridge and no linearization state to defend at inspection. That is where much of the drift lives.
There is an important caveat, though. Same platform is not the same result. The number still moves with the instrument unit, the analyst, the consumable lot, the sample pretreatment, and the target sequence selected for amplification within the same product — published comparisons on AAV bear that out. A same-platform deployment is still a method transfer with comparative testing. What is avoided is a method change, not a method difference.
When does a gene therapy lab need to upgrade from a research ddPCR system to a high-throughput one?
Sample count alone is a poor trigger. Three better ones to watch for:
Stability. Release testing is bursty; stability is a fixed calendar that cannot be renegotiated. Three batches, three conditions, five timepoints, and two ddPCR attributes is 90 determinations per pull season, arriving whether or not capacity exists. That is usually what breaks a research instrument first.
Multiplexing. Once two- or three-dimensional linkage genome integrity is needed alongside residual plasmid and a reference target, five or six channels are required in one well. Running those as separate simplex assays consumes drug product that may not be available.
Continuity. From pivotal supply onward, a single instrument is a single point of failure for batch release. A three-week service call becomes a release delay.
For scale: a walkaway nanoplate system handles roughly 380 samples a day on one plate module and up to approximately 1,200 on an eight-plate configuration. If realistic weekly demand approaches a single day's capacity, the threshold has already been crossed.
How does ddPCR batch release testing change when manufacturing scale increases?
Dilution becomes the dominant error term. Drug substance at 10¹³ to 10¹⁴ vg/mL has to reach roughly 10² to 10⁵ copies per microliter in the well, which means five or six serial dilution steps. Each contributes its own imprecision, and it compounds. At small scale that is manageable. At commercial scale, automating dilution is the single largest precision gain available.
Sample numbers also grow faster than batch numbers, because sub-lots and in-process points multiply.
The upside is statistical. Below roughly 20 batches, a specification is an educated guess. Past that threshold, there is enough data to separate method variance from process variance — which is exactly what the target measurement uncertainty in the USP ⟨1220⟩ analytical target profile (ATP) framework is designed for.
What can digital PCR not tell you about an AAV batch?
It tells you that a short amplicon is present in a DNase-resistant compartment. That is all it tells you.
It does not confirm the genome is full length: a 100 base pair (bp) amplicon is amplified identically by an intact genome and a truncated one. Linkage assays help, but in one published dataset a conventional titer appeared normal while two-dimensional linkage showed only approximately 50 percent of genomes carried both ends.
It does not resolve empty, partial, and overpacked capsids — that requires capsid titer and a mass or sedimentation method.
It says nothing about infectivity. A genome-containing capsid that cannot uncoat reads exactly like one that transduces efficiently.
And it is silent on ITR integrity, sequence fidelity outside the amplicon, aggregation, VP ratio, and post-translational modifications.
What do scientists most commonly get wrong about analytical continuity between R&D and GMP?
The most common misconception is that continuity means the same instrument. It does not. Continuity is the same procedure, the same reference material, the same data treatment, and the same acceptance criteria — with a documented rationale wherever one of those changed.
The second mistake follows from the first: assuming that because the method did not change, nothing changed. The sample did — new formulation, new excipients, a concentration a log higher, a different container closure.
The third is the cheapest to fix and the most often skipped: not banking retained material and qualifying an internal reference standard early. That standard is what anchors every later method change to a common scale.
And people forget that R&D data becomes regulatory data retroactively. Notebook entries written years before anyone used the phrase "Investigational New Drug (IND)" can end up supporting the dossier.
- Lock M, et al. Characterization of a recombinant adeno-associated virus type 2 reference standard material. Hum Gene Ther. 2010;21(10):1273–1285.
- Ayuso E, et al. Manufacturing and characterization of a recombinant adeno-associated virus type 8 reference standard material. Hum Gene Ther. 2014;25(11):977–987.
- Duong T, Firmo M, Li C-T, et al. Three-dimensional linkage analysis with digital PCR for genome integrity and identity of recombinant adeno-associated virus. Sci Rep. 2025;15:2154.
The content presented in this article is intended for research use only and is not intended for use in diagnostic procedures. The views and opinions expressed in the article are those of the individual researcher and do not necessarily reflect the official policy, position, or claims of Bio-Rad Laboratories.
Any clinical applications, test validations, or performance claims discussed are the sole responsibility of the featured researcher or their institution and do not reflect the views or endorsements of Bio-Rad Laboratories.
For Research Use Only. Not for use in diagnostic procedures.